Horizontal versus Vertical furnace arrangement

Total Sulfur and Total Nitrogen analysis

Introduction

In the analytical market, there has always been an ongoing discussion about which furnace setup performs the best in combustion analyzers. Some analysts claim a horizontal furnace placed version gives you the best performance, others swear on a vertical furnace setup.

TE instruments has recently developed a new model analyzer, Xplorer-V, for the analysis of total sulfur and total nitrogen levels in hydrocarbon products with a boiling point between 25 and 520 °C. The Xplorer-V is equipped with (as the name suggests) a furnace that is placed vertically for the combustion of samples. The analyzer is equipped with a sulfur detector that uses UV-fluorescence and a nitrogen detector that is based on chemiluminescence.

The Xplorer-V was launched as an extension of the existing portfolio of the similar already available analyzer from TE Instruments, the Xplorer TN/TS. The Xplorer TN/TS is based on the same analyzer principles as the Xplorer-V, utilizes the same detectors but has one main difference; the furnace is placed horizontally for the benefit of a boat-inlet for solid applications.

Horizontal VS Vertical

During the development and introduction of the new Xplorer-V analyzer, a validation of the performance characteristics was performed and compared against the Xplorer TN/TS. Both types of analyzers were simultaneously validated, and the results were used to look at biases in the performance between both analyzers.

The Xplorer-V was expected to perform at least as well as the current analyzer Xplorer TN/TS, in terms of repeatability and reproducibility. Changes in the design of the Xplorer-V analyzer, like the absence of a separate liquid inlet module and the vertically placed furnace, suggest it should be better able to measure components with a final boiling point of ≥ 420 °C without compromising repeatability (RSD) and/or recovery.

In the Xplorer TN/TS the furnace is placed horizontal, and a separate liquids injection module is used to evaporate the sample. This liquids introduction module is set at a fixed temperature of 500 °C. Hence the maximum boiling temperature for liquid samples is set at 420 °C for the Xplorer TN/TS to be safe and to prevent accumulation of sample material in the module.

Liquids module / Boat inlet

In case samples need to be injected with higher boiling ranges on the Xplorer TN/TS, a boat inlet module needs to be utilized instead of the liquids module introduction. With a boat inlet introduction, the sample is injected into a quartz cup filled with quartz wool. The cup is transferred into the high-temperature furnace at a pre-defined speed and depth. This method is great for high boiling samples and difficult matrices, as de introduction is nicely controlled by the speed of the cup movement. A minor drawback however, is de analysis time due to the cooling step at the end of the analysis. Sample analysis with a liquid introduction module tends to be around 3 to 6 minutes, while a boat introduction typically comes with an analysis time of approximately 5 to 9 minutes.

Xplorer with liquids module Xplorer with boat inlet

Picture 1: Left – liquids introduction on Xplorer TN/TS with liquids module, right – liquids introduction on Xplorer TN/TS with boat inlet.

Prior to the combustion of samples, controlled sample evaporation is key to achieve the best result. In the case of the Xplorer TN/TS, the boat inlet or liquids module handles this. For the Xplorer-V, the sample evaporation is controlled by a pre-combustion inlet which is placed inside the combustion tube. After evaporation of the sample, the material is carried into the hot zone of the combustion tube by the inert carrier gas. With the collision of oxygen gas, combustion of the sample takes place.

In the Xplorer-V, the sample is injected into pre-combustion inlet where the sample fully evaporates under an inert gas stream at 650°C prior to entering the vertically placed combustion tube. In the combustion tube, the sample reacts with oxygen for full combustion. Due to this new design, hydrocarbons with a high boiling point have less tendency to deposit in the “colder” pre-combustion inlet. Together with the carrier gas flow the sample also enjoys the gravitational forces which assists the sample to flow into the hotter bottom part of the combustion tube.

As mentioned, both analyzers are combustion analyzers in which a small amount of sample, typically ≤100 µL, is injected and combusted. The exhaust stream of this oxidation process is analyzed for the presence of sulfur dioxide and nitric oxide with the use of UV-fluorescence and chemiluminescence measuring principle.

For the validation of the new analyzer, a side-by-side bias comparison is made between the Xplorer-V and the Xplorer TN/TS.

During the validation and comparison, the following required and non-required parameters have been analyzed: Limit of Detection (LOD),  repeatability, recovery, memory effect, and Linearity. In this condensed paper, we present the results for repeatability and recovery as main parameters for comparison. Other parameters reach out of the scope of the topic of this paper.

Apparatus and method settings

Xplorer-V NS                                                                              Xplorer TN/TS

Picture 2: Xplorer-V NS                                                                                        Picture 3: Xplorer TN/TS

Vertical combustion analyzer:
Xplorer-V NS version with standard combustion tube inlet and Xpro-Vtm combustion tube. Liquid autosampler integrated in analyzer.

Horizontal combustion analyzer:
Xplorer TN/TS version with Xpro Combustion Tube with dual stage filters and liquids introduction
Module with Archie Liquids autosampler.

  Xplorer-V NS Xplorer TN/TS
Temperatures (°C)  
Inlet module 650 500
Zone 1 650 1000
Zone 2 1000 1000
Gas flows (ml/min)  
Carrier 100 100
Oxygen 400 400
Sample introduction  
Sample volume (ml) 100 100
Injection speed (ml/s) 1,0 1,0

Calibration

For calibration of the analyzers we made calibration lines with dissolved Pyridine in Xylene (Total N) and dissolved Thiophene in Xylene (Total S) according to the following dilution scheme:

  Concentration Total N (ppm) Concentration Total S (ppm)
Standard 1 0 0
Standard 2 1 1
Standard 3 2.5 2.5
Standard 4 5.0 5.0
Standard 5 7.5 7.5
Standard 6 10 10
Standard 7 25 25
Standard 8 50 50
Standard 9 75 75
Standard 10 100 100

 

The results of the calibration are presented in the 2 graphs below:

picture 1 picture 2

Results

A range of 8 different sample matrices ranging from low boiling to high boiling hydrocarbon liquids and material originated from biobased products like Biodiesel and FAME have been injected on both analyzer systems.

Each sample was injected from 10 individual vials. The results from both the nitrogen and the sulfur results are displayed in table 1 and table 2. As mentioned, the focus of this study was to check the reproducibility and accuracy of the different samples to compare the performance.

Table 1 Sulfur results:

Sample (N = 10) boiling point (°C) Horizontal (mg/kg) RSD (%) Vertical (mg/kg) RSD (%)
Diesel 150 – 380 5,60 0,75 5,80 0,69
Gasoline 100 – 400 4,01 0,55 4,03 1,12
CRM 100 mg/kg 140 101,29 0,70 99,81 0,14
CRM 10 mg/kg 140 9,98 0,47 10,00 0,36
Biodiesel 350 0,42 3,20 0,38 3,71
Bio Ethanol 65 2,20 2,56 2,71 2,71
Fame 430 7,62 1,66 7,46 1,35
Crude oil (Diluted) 260 – 280 26,42 0,19 26,90 2,09
VGO (Diluted) 425 – 510 151,81 0,32 150,17 0,70

 

Table 2 Nitrogen results:

Sample (N = 10) boiling point (°C) * Horizontal (mg/kg) RSD (%) Vertical (mg/kg) RSD (%)
Diesel 150 – 380 91,24 0,62 92,01 0,22
Gasoline 100 – 400 6,51 1,43 6,61 1,68
CRM 100 mg/kg 140 101,82 0,32 100,48 0,25
CRM 10 mg/kg 140 10,08 0,39 10,26 0,40
Biodiesel 350 1,73 3,29 1,88 2,86
Bio Ethanol 65 1,93 2,82 1,92 1,43
Fame 430 16,88 0,82 17,13 0,96
Crude oil (Diluted) 260 – 280 2,91 1,90 3,14 2,08
VGO (Diluted) 425 – 510 58,10 0,84 59,56 1,03

*For some samples, typical boiling ranges are reported, the exact boiling values were not available during time of reporting.

A vertically placed furnace gives a different peak shape compared to horizontal analyzer units. Peaks seem to reach an equilibrium in a faster way once reaching the detector chambers after combustion and keeps a steadily filled detector room. This is different compared to the peak behavior, which is seen in the Xplorer TN/TS analyzer.

To show the difference, below two recordings are presented from a Xplorer-V compared to the Xplorer TN/TS.  The flat spot you see in the left picture is not because we reached a detector max. It is because we get a continuous supply of analytes in the detector chambers and already reached an equilibrium.

peak Xplorer-V NS     peak Xplorer TN/TS

Picture 4: peak comparison of a 100 ppm CRM standard for Nitrogen and Sulfur, left Xplorer-V, right Xplorer TN/TS

Last notice on the results: The crude oil samples and VGO samples needed a small dilution to compensate for viscosity and to prevent coagulation of the material. Further on, for the VGO samples and crude oil, the boat introduction was required on the Xplorer TN/TS to prevent accumulation or blocking of the sample material on the inlet.

Conclusion Horizontal versus Vertical

From the tests run on both analyzer systems, it is hard to appoint a clear winner in our laboratories. Some samples run a little better on the vertical system, some on the horizontal system. In both situations, it is clearly seen that all different materials can be run on either one of these systems. It seems like it all comes back again to personal preferences and experience or situation-dependant.

If we were to put a sticker on both systems, we suggest the following:

  • Use a vertical analyzer system for performing daily analysis on liquid and gasses (+LPG) for samples ranging from a broad range of matrices, simple and steady with guaranteed high uptimes.
  • Use a horizontal analyzer for performing analysis on liquids, solids, and gas (+LPG) with samples from a broad range of matrices but respecting boiling ranges from the samples. The Xplorer TN/TS analyzer is more flexible because different liquid autosamplers are available, and solids can also be introduced with a dedicated autosampler.