The analysis of samples contained in pressurized cylinders is often considered to be challenging. Related sampling systems must be able to handle a wide range of sample matrices contained in cylinders at varying pressure levels. TE Instruments developed the GLS Gas & LPG autosampler to provide a safe and reliable solution. This for the analysis of Total Sulfur, Total Chloride, and speciated Halogens in Gaseous Hydrocarbons and Liquefied Petroleum Gases. The GLS stands out from other sampling systems with its safety features and ease of use.
The Gas & LPG autosampler functions as sample inlet system and provides a gas sampling loop and a LPG sampling loop with expansion chamber. The sampling system is swept by carrier gas (Ar or He) at pre-defined rate, which transfers the sample into the oxidation zone of the elemental combustion analyzer. This procedure is fully compliant with all relevant test methods. Such as the ASTM D3246, ASTM D6667, ASTM D7551, ASTM D7994, UOP 910, and UOP 1001. The GLS can be configured with different quick connects, sample loops, pressure modules, and connection methods. This creates the perfect settings for your application requirements.
Images
Safety is a critical topic related to Gas & LPG sampling due to the flammable nature of these matrices. Practically seen, the main fire risks arise from disconnecting an ‘open’ cylinder and having an incorrect connection between the cylinder and the autosampler. The GLS has several features to reduce safety risks related to Gas & LPG sampling:
• Automatic Safety Lock on Inlet Ports – Prevents the sample cylinder from being removed as long as pressure is detected in the connection between the GLS and the pressurized cylinder. When the cylinder is closed, the pressure will go down, and the safety lock is deactivated.
• Internal Gas leakage Sensor – Automatically aborts all activity when a small hydrocarbon spill is detected and warns the operator.
• Standard Supplied with female quick connector including auto-close valve – Minimize the risk of leakage. Enable a direct, leak-proof connection between the autosampler and the sample cylinder. The Auto close off valve provides an automatic flow lock if the connector is disconnected. By mounting a male quick connector with auto-close valve functionality, a double-ended shutoff connection is established.
• Real-time Software Notifications – Warn the operator about potential hazards and refer to SOP’s for LPG analysis.
Videos
Product categories
International Standards
EN 17178, GB/T 11061, UOP 936, UOP 910, GB/T 11060.4, ASTM D3246, ASTM D6667, ASTM D7551, IS 4576, JIS K 2240